Model 8060 Function Verifier
 |
The Model 8060 Function Verifier is a portable unit designed to be used in place of the conductivity cell to simulate conductivity ranges. The Model 8060 will verify proper operation of the conductivity, resistivity, total dissolved solids (TDS), temperature and salinity functions of the conductivity instrument. Easy to use, and requires no batteries, line power or maintenance. |
SPECIFICATIONS:
| Case: |
Off white, plastic ABS enclosure with sloping front panel |
| Dimensions: |
W= 7.49" D= 5.44" H=1.98/1.26" |
| Cable: |
One meter with 8 pin circular din (Male) connector |
| Weight: |
13.2 ounces |
| Env. Limits |
Temperature 5° C to 45° C |
| |
Humidity 10% to 90% relative non condensing |
| Conductivity Range |
Resistivity |
TDS |
Accuracy |
| |
|
|
|
| A 2.000 µS |
.500 M ohms |
1.250 PPM |
± .1% |
| B 20.00 µS |
50 K ohms |
12.50 PPM |
± .1% |
| C 200.0 µS |
5 K ohms |
125.0 PPM |
± .1% |
| D 2.000 mS |
N/A |
1.250 PPT |
± .1% |
| E 20.00 mS |
N/A |
12.50 PPT |
± .1% |
| F 200.0 mS |
N/A |
125.0 PPT |
± .1% |
| Salinity: |
35.0 Salinity Units at 15° C |
Accuracy = ± .1% |
| Temperature: |
15° C, 25° C, and 35° C |
Accuracy = ± .1% |
Model 8060 can be used with the following Conductivity instruments:
Model 3082 series, 3084, 4081, and Model 4083 multi-function EC meters.
Note: Design and specifications subject to change without notice.
Price: See General Price List |